    Table of Contents - (Show below) - (Hide below)

    FOREWORD
    INTRODUCTION
    1 Scope and object
    2 Normative references
    3 Terms and definitions
    4 General requirements
      4.1 PERFORMANCE REQUIREMENTS
      4.2 MEASURING ASSEMBLY
      4.3 Source types
          4.3.1 General
          4.3.2 Beta particle-emitting sources
          4.3.3 Low-energy-photon-emitting sources
      4.4 Quantity to be measured
      4.5 Reference and STANDARD TEST CONDITIONS
      4.6 General test conditions
          4.6.1 STANDARD TEST CONDITIONS
          4.6.2 STABILIZATION TIME
          4.6.3 Adjustments during test
          4.6.4 Batteries
      4.7 Constructional requirements as related to performance
          4.7.1 General
          4.7.2 Components
          4.7.3 Display
          4.7.4 Inserts
          4.7.5 STABILIZATION TIME
      4.8 Test of components
    5 Limits of performance characteristics
      5.1 Position of source in insert and repeatability
      5.2 USABLE LENGTH
      5.3 RESOLUTION OF THE DISPLAY
      5.4 STABILIZATION TIME
      5.5 LEAKAGE CURRENT
          5.5.1 In AIR KERMA STRENGTH measuring mode
          5.5.2 In charge measuring mode
      5.6 Stability
          5.6.1 Long term stability
          5.6.2 MANUFACTURER method to check long term stability
    6 LIMITS OF VARIATION for effects of influence quantities
      6.1 General
      6.2 IONIZATION CHAMBER - recombination losses
      6.3 Operating voltage
          6.3.1 Mains operated MEASURING ASSEMBLY
          6.3.2 Battery operated MEASURING ASSEMBLY
          6.3.3 Rechargeable MEASURING ASSEMBLY
      6.4 Air pressure
      6.5 Change of air pressure and EQUILIBRATION TIME
          of the radiation detector
          6.5.1 VENTED WELL TYPE IONIZATION CHAMBERS
          6.5.2 SEALED WELL TYPE IONIZATION CHAMBERS
      6.6 Temperature and humidity
      6.7 Length RESPONSE
      6.8 Electromagnetic immunity
    7 Marking
      7.1 WELL-TYPE IONIZATION CHAMBER ASSEMBLY
      7.2 MEASURING ASSEMBLY
    8 ACCOMPANYING DOCUMENTS
      8.1 General
      8.2 Use of the instrument
      8.3 Documentation
    Bibliography
    Index of defined terms

    Abstract - (Show below) - (Hide below)

    IEC 62467-1:2009 specifies the performance and some related constructional requirements of well-type ionization chambers and associated measurement apparatus, intended for the determination of a quantity, such as air kerma strength or reference air kerma rate in photon radiation fields or absorbed dose to water at a depth, in photon and beta radiation fields used in brachytherapy, after appropriate calibration for a given type of source. IEC 62467-1:2009 covers the techniques for the quantification of the quantity appropriate for the brachytherapy source under consideration. This quantity may be air kerma strength or reference air kerma rate at 1 m, or absorbed dose to water at a depth (e.g. 2 mm or 5 mm). Measurement of these quantities may be accomplished by a variety of well-type ionization chambers or systems currently available for this purpose. This standard applies to products intended for low dose rate, high dose rate, intravascular, both photon and beta, brachytherapy measurements. It does not apply to instruments for nuclear medicine applications. The application of the standard is limited to instruments that incorporate well-type ionization chambers as detectors. The intended use is the measurement of the output of radioactive, encapsulated sources for intracavitary (insertion into body cavities) or interstitial (insertion into body tissue) applications. The object of IEC 62467-1:2009 is
    a) to establish requirements for a satisfactory level of performance for well-type chamber systems, and
    b) to standardize the methods for the determination of compliance with this level of performance.
    IEC 62467-1:2009 is not concerned with the safety aspects of well-type chamber systems. The well-type chamber systems covered by this standard are not intended for use in patient environment. The electrical safety of well-type chamber systems is covered in IEC 61010-1. The operation of the electrometer measuring system is covered in IEC 60731.

    General Product Information - (Show below) - (Hide below)

    Committee TC 62/SC 62C
    Development Note Stability Date: 2025. (12/2017)
    Document Type Standard
    Publisher International Electrotechnical Committee
    Status Current

    Standards Referencing This Book - (Show below) - (Hide below)

    IEC 60050-394:2007 International Electrotechnical Vocabulary (IEV) - Part 394: Nuclear instrumentation - Instruments, systems, equipment and detectors
    IEC 60731:2011+AMD1:2016 CSV Medical electrical equipment - Dosimeters with ionization chambers as used in radiotherapy
    IEC 61187:1993 Electrical and electronic measuring equipment - Documentation
    IEC 61674:2012 Medical electrical equipment - Dosimeters with ionization chambers and/or semiconductor detectors as used in X-ray diagnostic imaging
    IEC 60601-1:2005+AMD1:2012 CSV Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
    IEC TR 60788:2004 Medical electrical equipment - Glossary of defined terms
    ISO/IEC Guide 99:2007 International vocabulary of metrology Basic and general concepts and associated terms (VIM)
    IEC 60580:2000 Medical electrical equipment - Dose area product meters
    IEC 60601-1-3:2008+AMD1:2013 CSV Medical electrical equipment - Part 1-3: General requirements forbasic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
    IEC 61676:2002+AMD1:2008 CSV Medical electrical equipment - Dosimetric instruments used for non-invasive measurement of X-ray tube voltage in diagnostic radiology
    IEC 61010-1:2010+AMD1:2016 CSV Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
    IEC 60050-393:2003 International Electrotechnical Vocabulary (IEV) - Part 393: Nuclear instrumentation - Physical phenomena and basic concepts
    • Access your standards online with a subscription

      Features

      • Simple online access to standards, technical information and regulations
      • Critical updates of standards and customisable alerts and notifications
      • Multi - user online standards collection: secure, flexibile and cost effective