• There are no items in your cart

ASTM C 1590 : 2004 : R2009

Superseded

Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by

Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry

Available format(s)

Hardcopy , PDF

Superseded date

11-11-2014

Language(s)

English

Published date

01-06-2009

€67.30
Excluding VAT

CONTAINED IN VOL. 12.01, 2015 Gives guidance for an alternate linear calibration for the determination of selected actinide isotopes in appropriately prepared aqueous solutions by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS).

Committee
C 26
DocumentType
Standard Practice
Pages
6
ProductNote
Reconfirmed 2009
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

1.1 This practice provides guidance for an alternate linear calibration for the determination of selected actinide isotopes in appropriately prepared aqueous solutions by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS). This alternate calibration is mass bias adjusted using thorium-232 (232Th) and uranium-238 (238U) standards. One of the benefits of this standard practice is the ability to calibrate for the analysis of highly radioactive actinides using calibration standards at much lower specific activities. Environmental laboratories may find this standard practice useful if facilities are not available to handle the highly radioactive standards of the individual actinides of interest.

1.2 The instrument response for a series of determinations of known concentration of 232Th and 238U defines the mass versus response relationship. For each standard concentration, the slope of the line defined by 232Th and 238U is used to derive linear calibration curves for each mass of interest using interference equations. The mass bias corrected calibration curves, although generated from interference equations, are specific to the instrument operating parameters and tuning in effect at the time of data acquisition. Because interference equations are part of the normal ICP-MS manufacturer's software package, this calibration methodology is widely applicable.

1.3 For this standard practice, the actinide atomic mass range that has been studied is from amu 232244. Guidance for an extended range of amu 228248 is given in this practice.

1.4 Using this practice, analyte concentrations are reported at each amu and not by element total (that is, 239Pu versus plutonium).

1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM C 1411 : 2014 : REDLINE Standard Practice for The Ion Exchange Separation of Uranium and Plutonium Prior to Isotopic Analysis
ASTM C 1168 : 2015 : REDLINE Standard Practice for Preparation and Dissolution of Plutonium Materials for Analysis

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.