Standard Test Method for On-Line pH Measurement of Water of Low Conductivity

Available format(s)

Hardcopy , PDF

Superseded date

11-11-2014

Superseded by

ASTM D 5128 : 2014

Language(s)

English

Published date

01-10-2009

€74.48
Excluding VAT

Committee
D 19
DocumentType
Test Method
Pages
11
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

1.1 This test method covers the precise on-line determination of pH in water samples of conductivity lower than 100 μS/cm (see Table 1 and Table 2) over the pH range of 3 to 11 (see Fig. 1), under field operating conditions, utilizing a sealed, non-refillable, reference electrode. pH measurements of water of low conductivity are problematical for conventional pH electrodes, methods, and related measurement apparatus.

1.2 This test method includes the procedures and equipment required for the continuous pH measurement of low conductivity water sample streams including the requirements for the control of sample stream pressure, flow rate, and temperature. For off-line pH measurements in low conductivity samples, refer to Test Method D 5464.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

TABLE 1 Calculated Conductivity and pH Values at 25°C of Low Concentrations of NaOH in Pure Water A

Note 1—This table tabulates the theoretical conductivity and pH values of low levels of NaOH in pure water as calculated from available thermodynamic data.

Note 2—To illustrate the high sensitivity of the sample pH at these low concentrations to contaminants, the last column lists errors that would result if the sample were contaminated with an additional 1 mg/L through sample or equipment handling errors.

Sample
Concentration,
mg/L
Sample
Conductivity,
μS/cm
Sample
pH
Δ pH Error from Additional 1 mg/L NaOH Contaminate
0.001 0.055 7.05Δ 2.35
0.010 0.082 7.45 Δ 1.95
0.100 0.625 8.40 Δ 1.03
1.0 6.229 9.40 Δ 0.30
8.0 49.83010.30 Δ 0.05

A Data courtesy of Ref (13). This data developed from algorithms originally published in Ref (14).

TABLE 2 Calculated Conductivity and pH Values at 25°C of Low Concentrations of HCl in Pure Water A

Note 1—This table tabulates the theoretical conductivity and pH values of low levels of HCl in pure water as calculated from available thermodynamic data

Note 2—To illustrate the high sensitivity of the sample pH at these low concentrations to contaminants, the last column lists errors that would result if the sample were contaminated with an additional 1 mg/L through sample or equipment handling errors.

Sample
Concentration,
mg/L
Sample
Conductivity,
μS/cm
Sample
pH
Δ pH Error from Additional 1 mg/L HCl Contaminate
0.001 0.0606.94Δ2.38
0.010 0.134 6.51 Δ 1.95
0.100 1.166 5.56 Δ 1.03
1.0 11.645 4.56 Δ 0.30
8.0 93.163 3.66 Δ 0.05

A Data courtesy of Ref (13). This data developed from algorithms originally published in Ref (14).


FIG. 1 Restrictions Imposed by the Conductivity pH Relationship

ASTM D 1193 : 2006 : R2018 Standard Specification for Reagent Water
ASTM D 1293 : 2018 Standard Test Methods for pH of Water
ASTM D 5464 : 2016 Standard Test Method for pH Measurement of Water of Low Conductivity
ASTM E 2635 : 2014 Standard Practice for Water Conservation in Buildings Through In-Situ Water Reclamation
ASTM D 6569 : 2014 Standard Test Method for On-Line Measurement of pH (Withdrawn 2023)

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.