ASTM D 7598 : 2009 : EDT 1
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
Standard Test Method for Determination of Thiodiglycol in Water by Single Reaction Monitoring Liquid Chromatography/Tandem Mass Spectrometry
Hardcopy , PDF
11-11-2014
English
01-12-2009
Committee |
D 19
|
DocumentType |
Test Method
|
Pages |
8
|
PublisherName |
American Society for Testing and Materials
|
Status |
Superseded
|
SupersededBy |
1.1 This procedure covers the determination of thiodiglycol (TDG) in surface water by direct injection using liquid chromatography (LC) and detected with tandem mass spectrometry (MS/MS). TDG is qualitatively and quantitatively determined by this method. This method adheres to single reaction monitoring (SRM) mass spectrometry.
1.2 This test method has been developed by US EPA Region 5 Chicago Regional Laboratory (CRL).
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 The Detection Verification Level (DVL) and Reporting Range for TDG are listed in Table 1.
1.4.1 The DVL is required to be at a concentration at least 3 times below the reporting limit (RL) and have a signal/noise ratio greater than 3:1. Fig. 1 displays the signal/noise ratio at the DVL.
1.4.2 The RL is the concentration of the level 1 calibration standard as shown in Table 2. The reporting limit for this method is 100 μg/L.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Analyte | DVL (μg/L) | Reporting Range (μg/L) |
Thiodiglycol | 20 | 100–10 000 |
Analyte/Surrogate | LV 1 | LV 2 | LV 3 | LV 4 | LV 5 | LV 6 | LV 7 |
Thiodiglycol | 100 | 250 | 500 | 1000 | 2500 | 5000 | 10 000 |
3,3’-Thiodipropanol | 100 | 250 | 500 | 1000 | 2500 | 5000 | 10 000 |
FIG. 1 Example SRM Chromatograms Signal/Noise at Detection Verification Level