ASTM E 839 : 2005
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
Standard Test Methods for Sheathed Thermocouples and Sheathed Thermocouple Material
Hardcopy , PDF
11-11-2014
English
01-11-2005
Committee |
E 20
|
DocumentType |
Test Method
|
Pages |
15
|
PublisherName |
American Society for Testing and Materials
|
Status |
Superseded
|
SupersededBy | |
Supersedes |
1.1 These are test methods for sheathed thermocouple material and assemblies.
1.2 The tests are intended to ensure quality control and to evaluate the suitability of sheathed thermocouple material or assemblies for specific applications. Some alternative test methods to obtain the same information are given, since in a given situation an alternative test method may be more practical. Service conditions are widely variable and so it is unlikely that all the tests described will be appropriate for a given thermocouple application. A brief statement is made following each test description to indicate when it might be used.
1.3 The tests described herein include test methods to measure the following properties of sheathed thermocouple material and assemblies.
1.3.1 Insulation PropertiesCompaction-absorption method and tension method.
Thickness.
Resistance-room temperature and elevated temperature.
1.3.2 Sheath PropertiesIntegrity-water test (two test methods) and mass spectrometer.
Dimensions-length, diameter, and roundness.
Wall thickness.
Surface-gross visual, finish, defect detection by dye penetrant, and cold-lap detection by tension test.
Metallurgical structure.
Ductility-bend test and tension test.
1.3.3 Thermoelement PropertiesCalibration.
Homogeneity.
Drift.
Thermoelement diameter, roundness, and surface appearance.
Thermoelement spacing.
Thermoelement ductility.
Metallurgical structure.
1.3.4 Thermocouple Assembly PropertiesDimensions-length, diameter, and roundness.
Surface-gross visual, finish, reference end seal, and defect detection by dye penetrant.
Electrical-continuity, loop resistance, and connector polarity.
Radiographic inspection.
Thermoelement diameter.
Thermal response time.
Thermal cycle.
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.