• There are no items in your cart

ASTM F 1862/F1862M : 2024 : REDLINE

Current

Current

The latest, up-to-date edition.

Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

Available format(s)

PDF

Language(s)

English

Published date

11-03-2024

1.1This test method is used to evaluate the resistance of medical face masks to penetration by the impact of a small volume (~2 mL) of a high-velocity stream of synthetic blood. Medical face mask pass/fail determinations are based on visual detection of synthetic blood penetration.

1.2This test method does not apply to all forms or conditions of blood-borne pathogen exposure. Users of the test method must review modes for face exposure and assess the appropriateness of this test method for their specific application.

1.3This test method is primarily intended to address the performance of finished medical face masks. While this test method may also be used to assess performance of materials or certain material constructions used in medical face masks, it is important to note the performance of finished medical face masks may be impacted by the interaction of the materials used and how they have been assembled. Results can differ depending on testing a final finished medical face mask or materials taken from manufactured medical face masks.

1.4This test method does not address other factors with the potential to affect the overall protection offered by the medical face mask and its operation (such as filtration efficiency and pressure drop).

1.5This test method does not address breathability of the medical face mask materials or any other properties affecting the ease of breathing through the medical face mask. This test method evaluates medical face masks as an item of protective clothing. This test method does not evaluate the performance of medical face masks for airborne exposure pathways or in the prevention of the penetration of aerosolized body fluids deposited on the medical face mask.

1.6The values stated in SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in nonconformance with the standard. The pressure values stated in each system are not exact equivalents. However, as the corresponding velocities are within 1 % of each other (see X1.4.2), reporting of the results in either units is permitted.

1.7This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.8This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Committee
F 23
DocumentType
Redline
Pages
42
PublisherName
American Society for Testing and Materials
Status
Current
Supersedes

View more information
€100.77
Excluding VAT

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.