• There are no items in your cart

ASTM F 2131 : 2002 : R2012

Superseded

Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by

Standard Test Method for In Vitro Biological Activity of Recombinant Human Bone Morphogenetic Protein-2 (rhBMP-2) Using the W-20 Mouse Stromal Cell Line

Available format(s)

Hardcopy , PDF

Superseded date

04-01-2022

Superseded by

ASTM F 2131 : 2021

Language(s)

English

Published date

01-11-2012

€67.30
Excluding VAT

CONTAINED IN VOL. 13.01, 2015 Describes the method used and the calculation if results for the determination of the in vitro biological activity rhBMP-2 using the mouse stromal cell line W-20 clone 17 (W-20-17).

Committee
F 04
DocumentType
Test Method
Pages
6
ProductNote
Reconfirmed 2012
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy
Supersedes

1.1This test method describes the method used and the calculation of results for the determination of the in-vitro biological activity of rhBMP-2 using the mouse stromal cell line W-20 clone 17 (W-20-17). This clone was derived from bone marrow stromal cells of the W++ mouse strain.2

1.2This test method (assay) has been qualified and validated based upon the International Committee on Harmonization assay validation guidelines3 (with the exception of interlaboratory precision) for the assessment of the biological activity of rhBMP-2. The relevance of this in vitro test method to in vivo bone formation has also been studied. The measured response in the W-20 bioassay, alkaline phosphatase induction, has been correlated with the ectopic bone-forming capacity of rhBMP-2 in the in vivo Use Test (UT). rhBMP-2 that was partially or fully inactivated by targeted peracetic acid oxidation of the two methionines was used as a tool to compare the activities. Oxidation of rhBMP-2 with peracetic acid was shown to be specifically targeted to the methionines by peptide mapping and mass spectrometry. These methionines reside in a hydrophobic receptor binding pocket on rhBMP-2. Oxidized samples were compared alongside an incubation control and a native control. The 62, 87, 98, and 100 % oxidized samples had W-20 activity levels of 62, 20, 7, and 5 %, respectively. The incubation and native control samples maintained 100 % activity. Samples were evaluated in the UT and showed a similar effect of inactivation on bone-forming activity. The samples with 62 % and 20 % activity in the W-20 assay demonstrated reduced levels of bone formation, similar in level with the reduction in W-20 specific activity, relative to the incubation control. Little or no ectopic bone was formed in the 7 and 5 % active rhBMP-2 implants.

1.3Thus, modifications to the rhBMP-2 molecule in the receptor binding site decrease the activity in both the W-20 and UT assays. These data suggest that a single receptor binding domain on rhBMP-2 is responsible for both in-vitro and in-vivo activity and that the W-20 bioassay is a relevant predictor of the bone-forming activity of rhBMP-2.

1.4The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ASTM F 2529 : 2013 Standard Guide for <emph type="bdit">in vivo</emph> Evaluation of Osteoinductive Potential for Materials Containing Demineralized Bone (DBM)
ASTM F 2211 : 2013 Standard Classification for Tissue Engineered Medical Products (TEMPs)
ASTM F 2529 : 2013 : R2021 Standard Guide for  <emph type="ital"> in vivo</emph> Evaluation of Osteoinductive Potential for Materials Containing Demineralized Bone (DBM)
ASTM F 2211 : 2013 : R2021 Standard Classification for Tissue-Engineered Medical Products (TEMPs)

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.