ASTM F 2149 : 2024
Current
The latest, up-to-date edition.
Standard Test Method for Automated Analyses of Cells—the Electrical Sensing Zone Method of Enumerating and Sizing Single Cell Suspensions
Hardcopy , PDF
English
01-12-2024
Committee |
F 04
|
DocumentType |
Test Method
|
Pages |
5
|
PublisherName |
American Society for Testing and Materials
|
Status |
Current
|
Supersedes |
1.1This test method, provided the limitations are understood, covers a procedure for both the enumeration and measurement of size distribution of many cell types. The instrumentation allows for user-selectable cell size settings and is applicable to a wide range of cell types. The method works best for spherical cells, and may be less accurate if cells are not spherical, such as for discoid cells or budding yeast. The method is appropriate for suspension as well as adherent cell cultures (1).2 Results may be reported as number of cells per milliliter or total number of cells per volume of cell suspension analyzed. Size distribution may be expressed in cell diameter or volume.
1.2Cells commonly used in tissue-engineered medical products (2) are analyzed routinely. Examples are chondrocytes (3), fibroblasts (4), and keratinocytes (5). Szabo et al. used the method for both pancreatic islet number and volume measurements (6). In addition, instrumentation using the electrical sensing zone technology was used for both count and size distribution analyses of porcine hepatocytes placed into hollow fiber cartridge extracorporeal liver assist systems. In this study (7), and others (6, 8), the automated electrical sensing zone method was validated for precision when compared to the conventional visual cell counting under a microscope using a hemocytometer. Currently, it is not possible to validate cell counting devices for accuracy since there is not a way to produce a reference sample that has a known number of cells. The electrical sensing zone method shall be validated each time it is implemented in a new laboratory, it is used on a new cell type, or the cell counting procedure is modified.
1.3Electrical sensing zone instrumentation is manufactured by several companies and is based upon electrical impedance. For many years, this field of cell analysis was dominated by the “Coulter” counter, which uses the electrical sensing zone principle. However, additional instruments that use this technology are now on the market, and development of the method for cell analysis applications is ongoing (9-14). This test method for cell counting and sizing is based on the detection and measurement of changes in electrical resistance produced by a cell, suspended in a conductive liquid, traversing through a small aperture (see Fig. 1 (15)). When cells are suspended in a conductive liquid, phosphate-buffered saline for instance, they function as discrete insulators. When the cell suspension is drawn through a small cylindrical aperture, the passage of each cell changes the impedance of the electrical path between two submerged electrodes located on each side of the aperture. An electrical pulse, suitable for both counting and sizing, results from the passage of each cell through the aperture. The path through the aperture, in which the cell is detected, is known as the “electronic sensing zone.” This test method permits the selective counting of cells within narrow size distribution ranges by electronic selection of the generated pulses. While the number of pulses indicates cell count, the amplitude of the electrical pulse produced depends on the cell's volume. The baseline resistance between the electrodes is due to the resistance of the conductive liquid within the boundaries of the aperture. The presence of cells within the “electronic sensing zone” raises the resistance of the conductive pathway that depends on the volume of the cell. Analyses of the behavior of cells within the aperture demonstrates that the height of the pulse produced by the cell is the parameter that most nearly shows proportionality to the cell volume.
1.4Limitations are discussed as follows:
1.4.1Coincidence—Occasionally, more than a single cell traverses the aperture simultaneously. Only a single larger pulse, as opposed to two individual pulses, is generated. The result is a lower cell count and higher cell volume measurement. The frequency of coincidence is a statistically predictable function of cell concentration that is corrected by the instrument. This is called coincidence correction (8). This phenomenon may be reduced by using lower cell concentrations.
1.4.2Viability—Electrical sensing zone cell counting enumerates both viable and nonviable cells and can estimate percent viable cells with a high degree of accuracy. A separate test such as Trypan blue dye exclusion or fluorescence flow cytometry is required to validate quantification of live cells and dead cells (16).
1.4.3Cell Diameter—This is a function of the size range capability of the aperture size selected. Measurements may be made in the cell diameter range of 0.6 μm to 1200 μm. Setting the counting size range on the instrument can affect the test results, especially if the cell size has a large distribution, and should be carefully controlled to help achieve repeatability.
1.4.4Size Range of the Aperture—The size range for a single aperture is proportional to its diameter. The response has been found to depend linearly on diameter over a range from 2 % to 80 % of the diameter. However, the aperture tube may become prone to blockage at levels greater than 60 % of diameter. Therefore, the practical operating range of the aperture is considered to be 2 % to 60 % of the diameter.
1.4.5Humidity—10 % to 85 %.
1.4.6Temperature—10 °C to 35 °C.
1.4.7Electrolyte Solution—The diluent for cell suspension shall provide conductivity and have minimal effect on cell size. The electrolyte of choice is commonly phosphate-buffered saline.
1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM E 2805 : 2018 | Standard Practice for Measurement of the Biological Activity of Ricin |
ASTM F 2739 : 2019 | Standard Guide for Quantifying Cell Viability and Related Attributes within Biomaterial Scaffolds |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.