ASTM F 2723 : 2013
NA
Status of Standard is Unknown
Standard Test Method for Evaluating Mobile Bearing Knee Tibial Baseplate/Bearing Resistance to Dynamic Disassociation
English
01-02-2013
| Committee |
F 04
|
| DocumentType |
Test Method
|
| Pages |
5
|
| PublisherName |
American Society for Testing and Materials
|
| Status |
NA
|
| SupersededBy | |
| Supersedes |
1.1This test method describes a laboratory method for evaluating the potential for mobile bearing knee tibial baseplate/bearing disassociation under repeated forces.
1.2The test described is applicable to any bicompartmental mobile bearing knee with a bearing retention mechanism. With modification, the test can be applied to a unicompartmental mobile bearing knee with a bearing retention mechanism.
1.3Although the methodology described does not replicate all physiological force conditions, it is a means of in vitro comparison of mobile bearing knee designs and the strength of the bearing retention mechanism between the tibial baseplate and bearing components under the stated test conditions.
| ASTM F 2083 : 2012 | Standard Specification for Knee Replacement Prosthesis |