ASTM F 2789 : 201
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
Standard Guide for Mechanical and Functional Characterization of Nucleus Devices
Hardcopy , PDF
11-11-2014
English
01-06-2010
CONTAINED IN VOL. 13.01, 2015 Specifies various forms of nucleus replacement and nucleus augmentation devices.
Committee |
F 04
|
DocumentType |
Guide
|
Pages |
10
|
ProductNote |
Reconfirmed 2010
|
PublisherName |
American Society for Testing and Materials
|
Status |
Superseded
|
SupersededBy |
1.1 This guide describes various forms of nucleus replacement and nucleus augmentation devices. It further outlines the types of testing that are recommended in evaluating the performance of these devices.
1.2 Biocompatibility of the materials used in a nucleus replacement device is not addressed in this guide. However, users should investigate the biocompatibility of their device separately (see X1.1).
1.3 While it is understood that expulsion and endplate fractures represent documented clinical failures, this guide does not specifically address them, although some of the factors that relate to expulsion have been included (see X1.3).
1.4 Multiple tests are described in this guide; however, the user need not use them all. It is the responsibility of the user of this guide to determine which tests are appropriate for the devices being tested and their potential application. Some tests may not be applicable for all types of devices. Moreover, some nucleus devices may not be stable in all test configurations. However, this does not necessarily mean that the test methods described are unsuitable.
1.5 The science of nucleus device design is still very young and includes technology that is changing more quickly than this guide can be modified. Therefore, the user must carefully consider the applicability of this guide to the user’s particular device; the guide may not be appropriate for every device. For example, at the time of publication, this guide does not address the nucleus replacement and nucleus augmentation devices that are designed to be partially or completely resorbable in the body. However, some of the test recommended in this guide may be applicable to evaluate such devices. It has not been demonstrated that mechanical failure of nucleus devices is related to adverse clinical results. Therefore this standard should be used with care in evaluating proposed nucleus devices.
1.6 This guide is not intended to be a performance standard. It is the responsibility of the user of this guide to characterize the safety and effectiveness of the nucleus device under evaluation.
1.7 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. Angular measurements may be reported in either degrees or radians.
1.8 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
ASTM F 3292 : 2019 | Standard Practice for Inspection of Spinal Implants Undergoing Testing |
ASTM F 1877 : 2016 : REDLINE | Standard Practice for Characterization of Particles |
ASTM E 328 : 2013 : REDLINE | Standard Test Methods for Stress Relaxation Tests for Materials and Structures |
ASTM F 561 : 2013 | Standard Practice for Retrieval and Analysis of Medical Devices, and Associated Tissues and Fluids |
ISO 18192-1:2011 | Implants for surgery Wear of total intervertebral spinal disc prostheses Part 1: Loading and displacement parameters for wear testing and corresponding environmental conditions for test |
ASTM E 1823 : 2013-07 | TERMINOLOGY RELATING TO FATIGUE AND FRACTURE TESTING |
ASTM E 132 : 2017 : REDLINE | Standard Test Method for Poisson’s Ratio at Room Temperature |
ASTM E 111 : 2017 : REDLINE | Standard Test Method for Young’s Modulus, Tangent Modulus, and Chord Modulus |
ASTM F 1980 : 2016 : REDLINE | Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices |
ASTM D 2990 : 2017 : REDLINE | Standard Test Methods for Tensile, Compressive, and Flexural Creep and Creep-Rupture of Plastics |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.