ASTM F 3169 : 2016 : R2024
Current
The latest, up-to-date edition.
Standard Test Method for Leak Detection in Blister Packaging by Vacuum Deflection Method by Laser Measurement
Hardcopy , PDF
English
01-11-2024
Committee |
F 02
|
DocumentType |
Test Method
|
Pages |
6
|
PublisherName |
American Society for Testing and Materials
|
Status |
Current
|
Supersedes |
1.1Test Packages—This test method can be applied to non-porous blister packs sealed with flexible films such as those used in pharmaceutical packaging. Such blister packs typically consist of thermoformed polymer or cold formed aluminum trays that contain a number of individual blister pockets into which tablets or capsules are placed. The trays are then sealed with a polymer, paper-backed or foil-based flexible laminate lidding material.
1.2Leaks Detected—This test method detects leaks in blister packs by measuring the deflection of the blister pack surface in response to an applied vacuum. This deflection of the blister pack surface results from the difference in pressure between the gas inside the blister pack and the applied vacuum. Air loss from within a blister pocket as a result of a leak will alter this pressure differential causing a measureable variation in blister pocket deflection. This test method requires that the blister packs are held in appropriate tooling inside a suitable test chamber.
1.3Test Results—Test results are reported qualitatively (pass/fail). Appropriate acceptance criteria for deflection, height, and collapse values are established by comparing non-leaking packs with those containing defects of a known size. Suitably sized defects in the laminate, tray material, and seal can be detected using this test method. The sensitivity of this test method depends upon a range of factors including blister pocket headspace, blister pocket size, lidding material type, lidding material thickness, lidding material tension, printing, surface texture, test conditions, and the values selected for the pass/fail acceptance criteria. The ability of the test to detect 15 µm, 50 µm, and catastrophic sized holes in four blister pack designs was demonstrated in a study.
1.4The values stated in SI units are to be regarded as standard and no other units of measurement are included in this test method.
1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM F 2097 : 2023 | Standard Guide for Design and Evaluation of Primary Flexible Packaging for Medical Products |