ASTM F 3275 : 2019

Superseded

Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by

Standard Guide for Using a Force Tester to Evaluate Performance of a Brush Part Designed to Clean the Internal Channel of a Medical Device

Available format(s)

Hardcopy , PDF

Superseded date

21-01-2022

Superseded by

ASTM F 3275 : 2022

Language(s)

English

Published date

21-02-2019

€74.48
Excluding VAT

Brushes used to clean a medical device after clinical use play an important role in effective reprocessing.

Committee
F 04
DocumentType
Guide
Pages
12
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy

1.1Brushes used to clean a medical device after clinical use play an important role in effective reprocessing. This guide describes methods for characterizing, under prescribed laboratory conditions, the efficacy of brush parts designed to clean the internal channels of medical devices. The methods utilize a force tester to mechanically actuate a brush part within a channel: (1) Methods to measure, at an established speed, the force required to move a brush within a channel; (2) Methods utilize the same force tester and protocols to measure soil removal from a soiled tube, another indicator of performance.

1.2Inclusions:

1.2.1This guide describes objective, quantifiable, and reproducible methods for evaluating the cleaning characteristics of a brush part under prescribed laboratory conditions, with test methods that simulate the cleaning challenge of a defined target area(s) of a medical device. This also makes possible the comparison of one design of a brush part to another.

1.2.2In this guide, a brush part is one that is intended to be moved within a tube.

1.2.3Tubes used for testing described in this guide are cylindrical and uniform in diameter. The test methods describe may not apply to non-cylindrical tubes.

1.2.4By use of this guide, medical device manufacturers can characterize the brush part designed for cleaning their device.

1.2.5By use of this guide, manufacturers of cleaning brushes can evaluate and characterize the cleaning performance of their brushes for the target area(s) of medical device(s), including allowing a comparison with existing brush part designs offered on the market. Further, they are able to evaluate modifications to designs and construction that might improve performance.

1.2.6This information can also be shared with the users of the brushes (medical device reprocessors) to help them evaluate the performance of commercially available brushes.

1.3Exclusions:

1.3.1This guide does not assess potential damage that may be inflicted by the brush. For instance, brushes with rigid bristles (for example, stainless steel or other metals), or other abrasive materials, are more likely to damage medical devices than brushes with flexible bristles (for example, nylon) or more pliable materials. Potential damage from more abrasive materials should be assessed.

1.3.2This guide does not specify acceptance criteria, and the results will be dependent on the specific parameters (for example, test soil, drying time, channel inside diameter and material, and so forth) that are tested.

1.3.3This guide is not intended to constitute all steps required to conduct validation of cleaning instructions for a medical device, including the use of brushes for this purpose, but provides methods that may be part of a broader protocol to conduct a complete cleaning instructions validation.

1.3.4If a brush is intended to clean a specific device(s), cleaning validation shall include testing with that device(s).

1.4Units—The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.6This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM F 3487 : 2020 Standard Guide for Assessing the Service Life of a Brush Part Intended to Clean a Medical Device

ASTM F 3208 : 2017 Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices
ASTM F 3208 : 2019 Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices
ASTM F 3208 : 2018 Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.