ASTM F 746 : 2004 : R2009 : EDT 1
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
Standard Test Method for Pitting or Crevice Corrosion of Metallic Surgical Implant Materials
Hardcopy , PDF
11-11-2014
English
01-12-2009
CONTAINED IN VOL. 13.01, 2015 Defines the determination of resistance to either pitting or crevice corrosion of metals and alloys from which surgical implants will be produced.
Committee |
F 04
|
DocumentType |
Test Method
|
Pages |
6
|
ProductNote |
Reconfirmed EDT 1
|
PublisherName |
American Society for Testing and Materials
|
Status |
Superseded
|
SupersededBy | |
Supersedes |
1.1 This test method covers the determination of resistance to either pitting or crevice corrosion of metals and alloys from which surgical implants will be produced. It is a modified version of an established test and is used as a screening test to rank surgical implant alloys in order of their resistance to localized corrosion.
1.2 This test method applies only to passive metals and alloys. Nonpassive alloys (other than noble alloys) are susceptible to general corrosion and are not normally suitable for implant use.
1.3 This test method is intended for use as a laboratory screening test for metals and alloys which undergo pitting or crevice corrosion, or both.
1.4 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
ASTM F 86 : 2013 : REDLINE | Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants |
ASTM G 3 : 2014 | Standard Practice for Conventions Applicable to Electrochemical Measurements in Corrosion Testing |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.