Standard Practice for Evaluating Acute Systemic Toxicity of Material Extracts by Systemic Injection in the Mouse

Available format(s)

PDF , PDF Redline

Language(s)

English

Published date

01-06-2020

Superseded date

11-02-2026

€57.49
Excluding VAT

Committee
F 04
DocumentType
Redline
Pages
4
PublisherName
American Society for Testing and Materials
Status
Superseded
SupersededBy

1.1This practice covers a nonspecific, acute toxicity test used for detecting leachables from materials used in medical devices.

1.2The liquids injected into the mouse are those obtained by Practice F619 where the extraction vehicles are saline, vegetable oil, or other liquids simulating human body fluids.

1.3Two procedures are outlined: Method A for intravenous injection and Method B for intraperitoneal injection.

1.4This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 may provide guidance for the selection of appropriate methods for testing materials for a specific application.

1.5The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.7This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM F 619 : 2014 : REDLINE Standard Practice for Extraction of Medical Plastics