1.1This practice covers a reference method of direct contact cell culture testing which may be used in evaluating the cytotoxic potential of materials for use in the construction of medical materials and devices.
1.2This practice may be used either directly to evaluate materials or as a reference against which other cytotoxicity test methods may be compared.
1.3This is one of a series of reference test methods for the assessment of cytotoxic potential, employing different techniques.
1.4Assessment of cytotoxicity is one of several tests employed in determining the biological response to a material, as recommended in Practice F748.
1.5The L-929 cell line was chosen because it has a significant history of use in assays of this type. This is not intended to imply that its use is preferred; only that the L-929 is a well characterized, readily available, established cell line that has demonstrated reproducible results in several laboratories.
1.6The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.7This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.8This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM F 1027 : 1986 : R1995 : EDT 1 | Standard Practice for Assessment of Tissue and Cell Compatibility of Orofacial Prosthetic Materials and Devices |
ASTM F 1027 : 1986 : R2017 | Standard Practice for Assessment of Tissue and Cell Compatibility of Orofacial Prosthetic Materials and Devices (Withdrawn 2023) |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.