• Shopping Cart
    There are no items in your cart

BS 5724-3.1(1990) : 1990 AMD 10622

Superseded

Superseded

A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.

View Superseded by

MEDICAL ELECTRICAL EQUIPMENT - PARTICULAR REQUIREMENTS FOR PERFORMANCE - METHODS OF DECLARING FUNCTIONAL PERFORMANCE CHARACTERISTICS OF MEDICAL ELECTRON ACCELERATORS IN THE RANGE 1 MNEV TO 50 MEV

Superseded date

28-09-2001

Superseded by

BS EN 60976 : 2008

Published date

23-11-2012

Sorry this product is not available in your region.

Committees responsible
National foreword
Method
1. Scope and object
2. Terminology
3. General information to the USER
4. Standardized test conditions
5. DOSE MONITORING SYSTEM
6. Depth ABSORBED DOSE characteristics
7. Uniformity of RADIATION FIELDS
8. Indication of RADIATION FIELDS
9. Indication of the RADIATION BEAM AXIS
10. ISOCENTRE
11. Indication of distance along the RADIATION BEAM
     AXIS
12. Zero position of rotational scales
13. Congruence of opposed RADIATION FIELDS
14. Movements of the PATIENT table
Appendices
A. Terminology
B. Format for presentation of functional performance
     values
ZA. Other international publications quoted in this
     standard with the references of the relevant
     European publications
Tables
1. Conditions for testing reproducibility
2. Conditions for testing proportionality of the DOSE
3. Conditions for testing dependence of the DOSE
     MONITORING SYSTEM on equipment position
4. Conditions for testing dependence of the DOSE
     MONITORING SYSTEM on GANTRY rotation
5. Conditions for testing dependence on the shape of
     the RADIATION FIELD
6. Conditions for testing stability of calibration
     of the DOSE MONITORING SYSTEM
7. Conditions for testing stability of the DOSE
     MONITORING SYSTEM in MOVING BEAM RADIOTHERAPY
8. Conditions for testing depth dose characteristics
9. Conditions for testing depth dose characteristics
10. Conditions for testing stability of PENETRATIVE
     QUALITY OF ELECTRON RADIATION
11. Flattened area according to Figure 4
12. Conditions for testing flatness and symmetry of
     X-RAY FIELDS
13. Conditions for testing deviation of dose
     distribution of X-RAY FIELDS with angular position
14. Conditions for testing maximum ABSORBED DOSE ratio
     in the RADIATION FIELD
15. Conditions for testing WEDGE FILTER FACTORS
16. Conditions for testing WEDGE FILTER ANGLES
17. Conditions for testing flatness, symmetry, deviation
     of dose distribution with angular position, and
     maximum ABSORBED DOSE ratio of ELECTRON FIELDS
18. Conditions for film calibration
19. Conditions for testing the numerical and the LIGHT
     FIELD-indication
20. Conditions for testing reproducibility of X-RAY
     FIELDS
21. Conditions for testing the LIGHT FIELD-INDICATOR
     for ELECTRON RADIATION
22. Conditions for testing geometry of the BEAM LIMITING
     SYSTEM
23. Conditions for testing the indication of the
     RADIATION BEAM AXIS on entry to the PATIENT
24. Conditions for testing the indication of the
     RADIATION BEAM AXIS on exit from the PATIENT
25. Conditions for testing indication of the ISOCENTRE
26. Conditions for testing vertical movement of the
     table
27. Conditions for testing ISOCENTRIC rotation of the
     table
28. Conditions for testing the angulation of rotational
     axes of the table
29. Conditions for testing lateral rigidity of the
     table
Figures
1. The rotary GANTRY
2. The wall- or floor-mounted GANTRY
3. The ceiling-mounted GANTRY
4. Flattened area (shown hatched) within the RADIATION
     FIELD
5. Examples of profiles of ABSORBED DOSE along the
     major axes or the diagonal axes
6. Explanatory diagram for flatness of the ELECTRON
     FIELD
7. Explanatory diagram for the definition of WEDGE
     FILTER ANGLE
8. Untitled
9. Test 8.1.1

Covers MEDICAL ELECTRON ACCELERATORS used for therapy purposes in human medical practice. Applies to ELECTRON ACCELERATORS which deliver a RADIATION BEAM of either X or ELECTRON RADIATION with ENERGIES in the range 1 MeV to 50 MeV at maximum ABSORBED DOSE RATES between 0,001 Gy s to the power of minus 1 and 1 Gy s to the power of minus 1 at 1m from the RADIATION SOURCE and at NORMAL TREATMENT DISTANCES between 50 cm and 200 cm from the RADIATION SOURCE. Describes test procedures to be undertaken by the manufacturer at design and construction stage but does not specify acceptance tests to be performed after installation on site.

Committee
HCC/81
DevelopmentNote
Renumbered and superseded by BS EN 60976. (10/2001) Supersedes 85/53310 DC and 98/563592 DC. (01/2006)
DocumentType
Standard
PublisherName
British Standards Institution
Status
Superseded
SupersededBy
Supersedes

BS 5724-3.1:Supplement No. 1:1990 Medical electrical equipment. Particular requirements for performance Methods of declaring functional performance characteristics of medical electron accelerators in the range 1 MeV to 50 MeV. Guide to functional performance values

BS 5724-3.1:Supplement No. 1:1990 Medical electrical equipment. Particular requirements for performance Methods of declaring functional performance characteristics of medical electron accelerators in the range 1 MeV to 50 MeV. Guide to functional performance values
BS 5724-1:1979 Medical electrical equipment Specification for general safety requirements

Access your standards online with a subscription

Features

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.