BS 5724-3.12:1991
Current
The latest, up-to-date edition.
Medical electrical equipment. Particular requirements for performance Method of declaring parameters for lung ventilators
Hardcopy , PDF
English
29-11-1991
Committees responsible
Foreword
Section 1. General
1 Scope
2 Definitions
Section 2. Test and operating conditions
3 Test conditions
3.1 General
3.2 Accuracy of measurement
3.3 Test apparatus
3.4 General test procedures, except for the pre-test
running described in 5.2
3.5 Reporting of test results in the ACCOMPANYING
DOCUMENTS
Section 3. Declaration and measurement of parameters
4 Performance parameters
5 Basic test procedures
5.1 General
5.2 Pre-test running
5.3 Waveform performance test
5.4 Expiratory resistance test (under CMV)
5.5 Resistance test (under spontaneous ventilation
conditions)
5.6 Volume performance test
5.7 I:E RATIO performance test
5.8 INSPIRATORY TIME performance test
5.9 Rebreathing test
5.10 Actions-of-controls test
5.11 MMV performance test
5.12 Synchronization test
5.13 Sequence of TRIGGERING sensitivity and delay tests
5.14 TRIGGERING-FREQUENCY range test
5.15 Over-maximum-FREQUENCY TRIGGERING waveform test
5.16 TRIGGERING waveform recording
6 LUNG VENTILATORS providing CMV
6.1 General
6.2 Waveform performance
6.3 Expiratory resistance
6.4 Volume performance
6.5 I:E RATIO performance
6.6 Rebreathing
6.7 Additional tests for LUNG VENTILATORS with
mechanisms for adjusting end-expiratory pressure
6.8 Actions of controls
6.9 Additional tests for LUNG VENTILATORS with
mechanisms for triggered CMV
7 LUNG VENTILATORS allowing spontaneous ventilation
which is either unsupplemented or supplemented
only by PRESSURE ASSISTANCE
7.1 General
7.2 Waveform performance
7.3 Resistance
7.4 Rebreathing
7.5 Additional tests for LUNG VENTILATORS with means
for the production of CPAP or CNAP
7.6 Additional tests for LUNG VENTILATORS with
adjustable mechanisms for assisting or hindering
spontaneous ventilation
8 LUNG VENTILATORS providing IMV
8.1 General
8.2 Tests with simulated spontaneous breaths only
8.3 Tests with mandatory breaths only
8.4 Tests with both mandatory and spontaneous breaths
9 LUNG VENTILATORS providing MMV
9.1 General
9.2 Tests with simulated spontaneous breaths only
9.3 Tests with mandatory breaths only
9.4 Tests with both intermittent mandatory and
spontaneous breaths or with pressure-assisted
mandatory breaths
Appendices
A Respiratory data on normal children
B Modes of operation
C Bibliography
Tables
1 'High' and 'low' flows for resistance tests
2 Required accuracies
3 Values corresponding to unit distance in displays
of waveform performance tests
4 Static characteristics of the rebreathing test
apparatus
5 Settings of the LUNG VENTILATOR and the active
part of the rebreathing test apparatus for the
rebreathing tests
6 Typical presentation of results of an actions-of-
controls test
7 Intervals corresponding to a given length of axis
in plots of the spontaneous/mandatory interaction
test
8 Settings of the test apparatus for a sequence of
TRIGGERING sensitivity and delay tests
9 Conditions for individual TRIGGERING sensitivity
and delay tests
10 Typical presentation of results of a sequence of
TRIGGERING sensitivity and delay tests
11 Conditions for the waveform performance test for
LUNG VENTILATORS providing CMV
12 Conditions for volume performance test in CMV
13 Conditions for I:E RATIO performance tests
14 Settings of the LUNG VENTILATOR for a sequence of
TRIGGERING sensitivity and delay tests in
triggered CMV and in SIMV
15 Settings of the LUNG VENTILATOR for the
TRIGGERING-FREQUENCY range test in triggered CMV
16 Typical presentation of results of a TRIGGERING-
FREQUENCY range test in CMV
17 Conditions for the waveform performance test for
LUNG VENTILATORS allowing spontaneous ventilation
18 Typical presentation of results of a TRIGGERING-
FREQUENCY range test in PRESSURE ASSISTANCE
19 Conditions for test 1 of waveform performance
tests in IMV (DELIVERED VOLUME and INSPIRATORY
TIME only) and MMV by IMV using only mandatory
breaths
20 Conditions for volume performance tests in IMV
21 Conditions for INSPIRATORY TIME tests in IMV
22 Conditions for waveform performance tests with
interaction of spontaneous breathing and MMV
23 Conditions for synchronization test recordings in
SIMV
24 Conditions for waveform performance tests with
interaction of spontaneous breathing and MMV
25 Conditions for MMV performance tests
26 Conditions for synchronization test recordings in
MMV by SIMV
27 Respiratory data on normal children
NUMEROUS FIGURES
Includes test methods to be used to measure parameters.
Committee |
CH/121/5
|
DevelopmentNote |
Supersedes 89/53567 DC and BS 3806(1964) (08/2005) Reviewed and confirmed by BSI, March 2013. (03/2013)
|
DocumentType |
Standard
|
Pages |
90
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
This standard describes a method of declaring the performance parameters to be stated in documents accompanying a lung ventilator, together with test methods to be used to measure the parameters. It does not cover parameters resulting from recent developments such as jet and very high frequency ventilation and oscillation. Motive power units for lung ventilators together with resuscitators, respiratory therapy ventilators, external body ventilators, rocking apparatus and electrostimulators are outside the scope of this standard. The object of this standard is to permit the direct comparison of lung ventilators of different manufacture, and to standardize methods for the determination of compliance with the manufacturer\'s declared performance. NOTE The titles of the publications referred to in this standard are listed on the inside back cover.
BS ISO 18562-1:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications Evaluation and testing within a risk management process |
ISO 18562-1:2017 | Biocompatibility evaluation of breathing gas pathways in healthcare applications — Part 1: Evaluation and testing within a risk management process |
15/30278530 DC : 0 | BS ISO 18562-1 - BIOCOMPATIBILITY EVALUATION OF BREATHING GAS PATHWAYS IN HEALTHCARE APPLICATIONS - PART 1: EVALUATION AND TESTING WITHIN A RISK MANAGEMENT PROCESS |
BS 5724-1(1989) : 1989 AMD 6715 | MEDICAL ELECTRICAL EQUIPMENT - GENERAL REQUIREMENTS FOR SAFETY |
BS 5724-2.13(1990) : 1990 | MEDICAL ELECTRICAL EQUIPMENT - PARTICULAR REQUIREMENTS FOR SAFETY - SPECIFICATION FOR ANAESTHETIC MACHINES |
BS 6015:1980 | Glossary of terms used in anaesthesiology |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.