BS EN 60118-14:1998
Current
The latest, up-to-date edition.
Hearing aids Specification of a digital interface device
Hardcopy , PDF
English
15-06-1998
INTRODUCTION
1 Scope
2 Normative references
3 Definitions
4 Symbols and abbreviations
5 Physical specifications of the interface
Annex ZA Normative references to international
publications with their corresponding European publications
Restricted to digital interface between control device and hearing aid with parameters set by direct electrical connection using a cable.
Committee |
EPL/29
|
DevelopmentNote |
Also numbered as IEC 60118-14. Supersedes 96/208900 DC. (09/2005)
|
DocumentType |
Standard
|
Pages |
12
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
This part of IEC60118 specifies the electrical and mechanical requirements for an interface device to allow a general-purpose control device such as a desktop computer to be connected to hearing aids, for the purpose of electrical setting of their operating parameters.
There is no specification of anything within the hearing aid, such as types of parameters and their signal processing characteristics, nor is there any specification of the parameter setting procedure — the method used to determine which settings are most appropriate to the person wearing the hearing aid. This standard is restricted to the digital interface between the control device and the hearing aid.
This standard provides a specification such that all hearing aids which can be programmed by the specified digital interface may have their parameters set by a single control device and interface device.
The specifications of this standard are for use in connection with hearing aids with parameters set by a direct electrical connection using a cable. They do not cover hearing aids with parameters controlled by radio frequency or infrared transmission, or any other “wireless” method of control.
Standards | Relationship |
DIN EN 60118-14:1998-11 | Identical |
NF EN 60118-14 : 1998 | Identical |
EN 60118-14:1998 | Identical |
NBN EN 60118-14 : 1998 | Identical |
I.S. EN 60118-14:1998 | Identical |
SN EN 60118-14 : 1998 | Identical |
UNE-EN 60118-14:2000 | Identical |
EN 60601-1-1:2001 | Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems |
EN 60118-12:1996 | Hearing aids - Part 12: Dimensions of electrical connector systems |
IEC 61076-4-105:1995 | Connectors with assessed quality, for use in d.c., low frequency analogue and in digital high-speed data applications - Part 4: Printed board connectors - Section 105 : Detail specification for 9 mm circular connector with 3 to 8 contacts for use in a wide range of applications including the telecommunication . |
IEC 60118-12:1996 | Hearing aids - Part 12: Dimensions of electrical connector systems |
IEC 60601-1-1:2000 | Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems |