1 Scope
2 Normative references
3 Terms and definitions
4 Methods to determine the ATTENUATION RATIO
5 Determination of ATTENUATION properties
6 Statement of compliance
Annex A (informative) - Tables of ATTENUATION RATIOS,
BUILD-UP FACTORS and first HALF-VALUE LAYERS
Bibliography
Index of defined terms used in this International Standard
Annex ZA (normative) - Normative references to
international publications with their corresponding
European publications
Pertains to materials in sheet form used for the manufacturing of PROTECTIVE DEVICES against X-RADIATION of RADIATION QUALITIES generated with X-RAY TUBE VOLTAGES up to 400 kV and gamma radiation emitted by radionuclides with photon energies up to 1,3 MeV.
Committee |
CH/62/2
|
DevelopmentNote |
Renumbers and supersedes BS 7817-1(1995). 2002 Version incorporates amendment 13968 to BS 7817-1(1995). (12/2002) 2002 Edition remains active and will be withdrawn on 11/06/2017. Supersedes 12/30242777 DC. (11/2014)
|
DocumentType |
Standard
|
Pages |
30
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
IEC 61331:2014 applies to materials in sheet form used for the manufacturing of protective devices against X-radiation of radiation qualities generated with X-ray tube voltages up to 400 kV and gamma radiation emitted by radionuclides with photon energies up to 1,3 MeV. This Part 1 is not intended to be applied to protective devices when these are to be checked for the presence of their attenuation properties before and after periods of use. This Part 1 specifies the methods of determining and indicating the attenuation properties of the materials. The attenuation properties are given in terms of:
- attenuation ratio;
- build-up factor; and
- attenuation equivalent together with, as appropriate, an indication of homogeneity and mass per unit area. Ways of stating values of attenuation properties in compliance with this part of the International Standard are included. Excluded from the scope of this International Standard are the methods for periodical checks of protective devices, particularly of protective clothing, the methods of determining attenuation by layers in the radiation beam, and the methods of determining attenuation for purposes of protection against ionizing radiation provided by walls and other parts of an installation. This second edition cancels and replaces the first edition of IEC 61331-1, published in 1994. It constitutes a technical revision. This second edition has been adapted to apply to the present technology. In particular, this second edition is consistently applicable to lead- and non-lead-containing materials. The essential changes and extensions are:
- the extension of the scope to cover photon-emitting radionuclides;
- improved methods to determine the attenuation ratio;
- the addition of the so-called inverse broad beam condition;
- the addition of a method to calculate the attenuation ratio of photon-emitting radionuclides;
- the definition of new standard X- and gamma radiation qualities used for testing;
- the addition of the so-called lead equivalent class; and
- tables of attenuation ratios, build-up factors and first half-value layers for the standard radiation qualities filtered with different thicknesses of lead.
Standards | Relationship |
UNE-EN 61331-1:2015 | Identical |
IEC 61331-1:2014 | Identical |
NF EN 61331-1 : 2015 | Identical |
DIN EN 61331-1:2016-09 | Identical |
NBN EN 61331-1 : 2014 | Identical |
EN 61331-1:2014 | Identical |
I.S. EN 61331-1:2014 | Identical |
EN 60601-1-3 : 2008 AMD 11 2016 | MEDICAL ELECTRICAL EQUIPMENT - PART 1-3: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: RADIATION PROTECTION IN DIAGNOSTIC X-RAY EQUIPMENT (IEC 60601-1-3:2008/A1:2013) |
EN 61331-3:2014 | Protective devices against diagnostic medical X-radiation - Part 3: Protective clothing, eyewear and protective patient shields |
IEC 60601-1:2005+AMD1:2012 CSV | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance |
IEC TR 60788:2004 | Medical electrical equipment - Glossary of defined terms |
IEC 60601-1-3:2008+AMD1:2013 CSV | Medical electrical equipment - Part 1-3: General requirements forbasic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment |
ISO 4037-1:1996 | X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 1: Radiation characteristics and production methods |
IEC 61331-3:2014 | Protective devices against diagnostic medical X-radiation - Part 3: Protective clothing, eyewear and protective patient shields |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.