This part of ISO11140 specifies the requirements for Class 2 indicators for Bowie and Dick-type air removal tests used to evaluate the effectiveness of air removal during the pre-vacuum phase of pre-vacuum steam sterilization cycles.
Additionally, this part of ISO11140 includes test methods and equipment used to meet these performance requirements.