BS ISO 16063-12:2002
Current
The latest, up-to-date edition.
Methods for the calibration of vibration and shock transducers Primary vibration calibration by the reciprocity method
Hardcopy , PDF
English
29-07-2002
Foreword
1 Scope
2 Normative references
3 Uncertainty of measurement
4 Symbols
5 Requirements for apparatus
6 Ambient conditions
7 Preferred amplitudes and frequencies
8 Procedure
9 Computation of sensitivity
Annex A (normative) - Calculation of uncertainty
Annex B (informative) - Application of the theory
of reciprocity to the calibration of
electromechanical transducers
Bibliography
Describes the instrumentation and procedures to be used for primary calibration of accelerometers using the reciprocity method and the SI system of units.
Committee |
GME/21
|
DevelopmentNote |
Supersedes BS 6955-20(1997) and 99/717571 DC (08/2002) Reviewed and confirmed by BSI, April 2008. (03/2008)
|
DocumentType |
Standard
|
Pages |
28
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
This part of ISO16063 specifies the instrumentation and procedures to be used for primary calibration of accelerometers using the reciprocity method and the SI system of units.
It is applicable to the calibration of rectilinear accelerometers over a frequency range of 40Hz to 5kHz and a frequency-dependent amplitude range of 10m/s2 to 100m/s2 and is based on the use of the coil of an electrodynamic vibrator as the reciprocal transducer.
Calibration of the sensitivity of a transducer can be obtained using this part of ISO16063 provided that the signal conditioner or amplifier used with the transducer during calibration has been adequately characterized. In order to achieve the uncertainties of measurement given in clause3, it has been assumed that the transducer has been calibrated in combination with its signal conditioner or amplifier (the combination of which in this part of ISO16063 is referred to as the “accelerometer”).
Standards | Relationship |
ISO 16063-12:2002 | Identical |
ISO 266:1997 | Acoustics — Preferred frequencies |
ISO 16063-1:1998 | Methods for the calibration of vibration and shock transducers Part 1: Basic concepts |