EN 14254:2004
Superseded
A superseded Standard is one, which is fully replaced by another Standard, which is a new edition of the same Standard.
View Superseded by
In vitro diagnostic medical devices - Single-use receptacles for the collection of specimens, other than blood, from humans
18-09-2021
30-06-2004
Foreword
1 Scope
2 Normative references
3 Terms and definitions
4 Materials
5 Nominal liquid capacity
6 Graduation and fill lines
6.1 Graduation lines
6.2 Minimum fill line
6.3 Maximum fill line
7 Design
8 Construction
9 Sterility and special microbiological states
10 Additives
11 Information supplied by the manufacturer
Annex A (normative) Tests for nominal capacity, graduation
lines and minimum free space for
non-evacuated specimen receptacles up
to and including 30 ml nominal capacity
A.1 Reagents and apparatus
A.2 Test Conditions
A.3 Test procedure
Annex B (normative) Tests for nominal capacity, graduation
lines and minimum free space, for
non-evacuated specimen receptacles of
greater than 30 ml nominal capacity
B.1 Reagents and apparatus
B.2 Test conditions
B.3 Test procedures
Annex C (normative) Tests for draw volume and minimum free
space for evacuated receptacles
C.1 Reagents and apparatus
C.2 Test conditions
C.3 Test procedure
Annex D (normative) Test for leakage from the closure of
a receptacle
D.1 Reagents and apparatus
D.2 Test procedure for non-evacuated receptacles intended
for storage above 0 degree C
D.3 Test procedure for non-evacuated receptacles intended
for storage at 0 degree C or below
D.4 Test procedure for evacuated receptacles
Annex E (normative) Test for the robustness of a receptacle
that is intended for centrifugation
E.1 Reagents and apparatus
E.2 Test conditions
E.3 Test procedure
Annex ZA (informative) Relationship between this European
Standard and the Essential
Requirements of EU Directive 98/79/EC
Bibliography
This standard specifies requirements and test methods for single-use evacuated and non evacuated receptacles, intended by their manufacturers, for the primary containment and preservation of specimens, other than blood specimens, derived from the human body, for the purposes of in vitro diagnostic examination.NOTE 1Requirements and test methods for evacuated and non evacuated single use venous blood specimen containers are specified in prEN 14820.NOTE 2While it is desirable that specimen receptacles should be designed to avoid spontaneous discharge of the contents, when being opened, this standard does not specify a test procedure for this because it has not been possible to devise an objective and reproducible test.This standard does not specify requirements for collection needles or needle holders or other accessories used in conjunction with specimen receptacles.
Committee |
CEN/TC 140
|
DevelopmentNote |
Supersedes PREN 14254 (10/2005)
|
DocumentType |
Standard
|
PublisherName |
Comite Europeen de Normalisation
|
Status |
Superseded
|
SupersededBy |
Standards | Relationship |
BS EN 14254:2004 | Identical |
NS EN 14254 : 1ED 2004 | Identical |
UNI EN 14254 : 2004 | Identical |
SN EN 14254 : 2004 | Identical |
PN EN 14254 : 2009 | Identical |
NEN EN 14254 : 2004 | Identical |
GOST R EN 14254 : 2010 | Identical |
NBN EN 14254 : 2004 | Identical |
UNE-EN 14254:2004 | Identical |
I.S. EN 14254:2004 | Identical |
NF EN 14254 : 2004 | Identical |
ONORM EN 14254 : 2004 | Identical |
DIN EN 14254:2004-09 | Identical |
BS EN 14820:2004 | Single-use containers for human venous blood specimen collection |
I.S. EN 14820:2004 | SINGLE-USE CONTAINERS FOR HUMAN VENOUS BLOOD SPECIMEN COLLECTION |
EN 14820:2004 | Single-use containers for human venous blood specimen collection |
DIN EN 14820:2004-11 | SINGLE-USE CONTAINERS FOR HUMAN VENOUS BLOOD SPECIMEN COLLECTION |
EN 550 : 1994 | STERILIZATION OF MEDICAL DEVICES - VALIDATION AND ROUTINE CONTROL OF ETHYLENE OXIDE STERILIZATION |
ISO 4788:2005 | Laboratory glassware — Graduated measuring cylinders |
EN 1041:2008+A1:2013 | Information supplied by the manufacturer of medical devices |
EN 552:1994/A2:2000 | STERILIZATION OF MEDICAL DEVICES - VALIDATION AND ROUTINE CONTROL OF STERILIZATION BY IRRADIATION |
EN 375 : 2001 | INFORMATION SUPPLIED BY THE MANUFACTURER WITH IN VITRO DIAGNOSTIC REAGENTS FOR PROFESSIONAL USE |
EN 829 : 1996 | IN VITRO DIAGNOSTIC SYSTEMS - TRANSPORT PACKAGES FOR MEDICAL AND BIOLOGICAL SPECIMENS - REQUIREMENTS, TESTS |
ISO 3696:1987 | Water for analytical laboratory use — Specification and test methods |
EN ISO 3696:1995 | Water for analytical laboratory use - Specification and test methods (ISO 3696:1987) |
ISO 7000:2014 | Graphical symbols for use on equipment Registered symbols |
EN 980:2008 | Symbols for use in the labelling of medical devices |
EN 14820:2004 | Single-use containers for human venous blood specimen collection |
EN 554 : 1994 | STERILIZATION OF MEDICAL DEVICES - VALIDATION AND ROUTINE CONTROL OF STERILIZATION BY MOIST HEAT |
ISO 15223:2000 | Medical devices Symbols to be used with medical device labels, labelling and information to be supplied |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.