ISO 17862:2022
Current
The latest, up-to-date edition.
Surface chemical analysis — Secondary ion mass spectrometry — Linearity of intensity scale in single ion counting time-of-flight mass analysers
Hardcopy , PDF , PDF 3 Users , PDF 5 Users , PDF 9 Users
English
23-09-2022
This document specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95% linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid.
DocumentType |
Standard
|
Pages |
18
|
PublisherName |
International Organization for Standardization
|
Status |
Current
|