MEDICAL ELECTRICAL EQUIPMENT - DOSE AREA PRODUCT METERS

Superseded date

01-11-2021

Superseded by

NF EN IEC 60580:2020

Published date

12-01-2013

Sorry this product is not available in your region.

AVANT-PROPOS
INTRODUCTION
1 Domaine d'application et objet
2 Références normatives
3 Terminologie et définitions
4 Prescriptions générales
  4.1 Prescriptions de performance
  4.2 DOMAINES UTILES minimaux de PRODUIT EXPOSITION-SURFACE
       et de DÉBIT DE PRODUIT EXPOSITION-SURFACE
  4.3 Plan de mesurage
  4.4 VALEURS DE RÉFÉRENCE et les CONDITIONS D'ESSAI
       NORMALISÉES
  4.5 Conditions générales d'essai
       4.5.1 CONDITIONS D'ESSAI NORMALISÉES
       4.5.2 Essai des composants
       4.5.3 TEMPS DE STABILISATION
       4.5.4 Réglages en cours d'essai
       4.5.5 Uniformité du champ de rayonnement
  4.6 Fluctuations statistiques
  4.7 Incertitude de la mesure
  4.8 Règles de construction liées aux performances
       4.8.1 Affichage
       4.8.2 Indication d'un défaut de tension de polarisation
       4.8.3 Dépassement de l'étendue de mesure
       4.8.4 Indication de réinitialisation ou autre état
             d'inactivité
       4.8.5 CHAMBRE D'IONISATION
  4.9 CONTROLEUR DE CONSTANCE
  4.10 Réglage
  4.11 Sécurité électrique
5 Limites des CARACTÉRISTIQUES DE PERFORMANCE SOUS CONDITIONS
  D'ESSAI NORMALISÉES
  5.1 ERREUR INTRINSEQUE RELATIVE
  5.2 Fonction d'alarme
  5.3 Répétabilité
  5.4 POUVOIR DE RÉSOLUTION de lecture
  5.5 TEMPS DE STABILISATION
  5.6 Réinitialisation des plages de PRODUIT EXPOSITION-SURFACE
  5.7 Dérive des VALEUR INDIQUÉES
  5.8 Stabilité à long terme
  5.9 TEMPS DE RÉPONSE
  5.10 Uniformité spatiale de la RÉPONSE
6 LIMITES DE VARIATION pour les effets des GRANDEURS D'INFLUENCE
  6.1 Dépendance de l'énergie de la RÉPONSE
  6.2 Dépendance des mesures de PRODUIT EXPOSITION-SURFACE du
       DÉBIT DE PRODUIT EXPOSITION-SURFACE
       6.2.1 ENSEMBLE DE MESURAGE
       6.2.2 CHAMBRE D'IONISATION - pertes de recombinaison
  6.3 DURÉE D'IRRADIATION
  6.4 Taille du champ
  6.5 Tension de fonctionnement
  6.6 Pression d'air
  6.7 Température et humidité
  6.8 Fluctuation de la densité de l'air dans la CHAMBRE
       D'IONISATION
  6.9 Compatibilité électromagnétique
       6.9.1 Généralités
       6.9.2 Décharges électrostatiques
       6.9.3 Champs électromagnétiques rayonnés
       6.9.4 Perturbations conduites induites par les salves
             et les hautes fréquences
       6.9.5 Ondes de choc
       6.9.6 Creux de tension, coupures brèves et VARIATIONS
             de tension
  6.10 INCERTITUDE NORMALE COMBINÉE
7 Marquage
  7.1 ENSEMBLE DE MESURAGE
  7.2 CHAMBRE D'IONISATION
8 DOCUMENTS D'ACCOMPAGNEMENT
Bibliographie
Index des termes définis
Annexe ZA (normative) Références normatives à d'autres
                      publications internationales avec les
                      publications européennes correspondantes

Defines performance and testing of DOSE AREA PRODUCT METERS with IONIZATION CHAMBERS for measurement of DOSE AREA PRODUCT and/or DOSE AREA PRODUCT RATE to which the PATIENT is exposed during MEDICAL RADIOLOGICAL EXAMINATIONS. Objectives are to: establish criteria for a satisfactory performance level for DOSE AREA PRODUCT METERS, and standardize the methods for determining compliance with this level of performance.

DevelopmentNote
Indice de classement: C74-103. PR NF NF EN 60580 December 2003. (12/2003)
DocumentType
Standard
PublisherName
Association Francaise de Normalisation
Status
Superseded
SupersededBy

Standards Relationship
DIN EN 60580:2001-12 Identical
BS EN 60580:2000 Identical
EN 60580:2000 Identical
I.S. EN 60580:2000 Identical
IEC 60580:2000 Identical
SN EN 60580 : 2000 Identical
NBN EN 60580 : 2001 Identical
UNE-EN 60580:2001 Identical

NF EN 60601-2-54 : 2009 AMD 1 2015 MEDICAL ELECTRICAL EQUIPMENT - PART 2-54: PARTICULAR REQUIREMENTS FOR THE BASIC SAFETY AND ESSENTIAL PERFORMANCE OF X-RAY EQUIPMENT FOR RADIOGRAPHY AND RADIOSCOPY

NF EN 61187 : 1995 ELECTRICAL AND ELECTRONIC MEASURING EQUIPMENT - DOCUMENTATION
IEC 60731:2011+AMD1:2016 CSV Medical electrical equipment - Dosimeters with ionization chambers as used in radiotherapy
IEC 60950:1999 Safety of information technology equipment
IEC 61187:1993 Electrical and electronic measuring equipment - Documentation
NF EN 61000 4-2 : 2009 ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 4-2: TESTING AND MEASUREMENT TECHNIQUES - ELECTROSTATIC DISCHARGE IMMUNITY TEST
IEC 60601-1:2005+AMD1:2012 CSV Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 61000-4-11:2004+AMD1:2017 CSV Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests
NF EN 61000-4-4 : 2013 ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 4-4: TESTING AND MEASUREMENT TECHNIQUES - ELECTRICAL FAST TRANSIENT/BURST IMMUNITY TEST
IEC TR 60788:2004 Medical electrical equipment - Glossary of defined terms
NFC 74 000 : 1989 MEDICAL RADIOLOGY - TERMINOLOGY
NF EN 60601-1-2 : 2016 MEDICAL ELECTRICAL EQUIPMENT - PART 1-2: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: ELECTROMAGNETIC DISTURBANCES - REQUIREMENTS AND TESTS
IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 61000-4-2:2008 Electromagnetic compatibility (EMC) - Part 4-2: Testing and measurement techniques - Electrostatic discharge immunity test
IEC 61000-4-3:2006+AMD1:2007+AMD2:2010 CSV Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test
NF EN 60731 : 2012 MEDICAL ELECTRICAL EQUIPMENT - DOSIMETERS WITH IONIZATION CHAMBERS AS USED IN RADIOTHERAPY
NF EN 61000 4-6 : 2014 ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 4-6: TESTING AND MEASUREMENT TECHNIQUES - IMMUNITY TO CONDUCTED DISTURBANCES, INDUCED BY RADIO-FREQUENCY FIELDS
HD 501 : 200S1 MEDICAL RADIOLOGY - TERMINOLOGY
NF EN 61000-4-11 : 2004 AMD 1 2017 ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 4-11: TESTING AND MEASUREMENT TECHNIQUES - VOLTAGE DIPS, SHORT INTERRUPTIONS AND VOLTAGE VARIATIONS IMMUNITY TESTS
IEC 60601-1-1:2000 Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems