• Shopping Cart
    There are no items in your cart

UNE-EN IEC 60601-2-20:2020

Current

Current

The latest, up-to-date edition.

Medical electrical equipment - Part 2-20: Particular requirements for the basic safety and essential performance of infant transport incubators (Endorsed by Asociación Española de Normalización in November of 2020.)

Available format(s)

Hardcopy , PDF

Published date

01-11-2020

€89.00
Excluding VAT

201.1 Scope, object and related standards

Clause 1 of the general standard applies, except as follows:

201.1.1 Scope

Replacement:

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of INFANT TRANSPORT INCUBATOR equipment, as defined in 201.3.211 of this standard, also referred to as ME EQUIPMENT.

If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.

HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of the general standard.

NOTE See also 4.2 of the general standard.

This particular standard specifies safety requirements for INFANT TRANSPORT INCUBATORS but alternate methods of compliance with a specific clause by demonstrating equivalent safety will not be judged as non compliant if the MANUFACTURER has demonstrated in his RISK MANAGEMENT FILE that the RISK presented by the HAZARD has been found to be of an acceptable level when weighed against the benefit of treatment from the device.

This particular standard does not apply to:

devices supplying heat via BLANKETS, PADS or MATTRESSES in medical use; for information see IEC 80601-2-35 [1]1);

INFANT INCUBATORS which are not INFANT TRANSPORT INCUBATOR; for information see IEC 60601-2-19 [2];

INFANT RADIANT WARMERS; for information, see IEC 60601-2-21 [3];

INFANT PHOTOTHERAPY; for information, see IEC 60601-2-50 [4].

201.1.2 Object

Replacement:

The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL PERFORMANCE requirements for INFANT TRANSPORT INCUBATORS as defined in 201.3.211, which minimize HAZARDS to the PATIENT and OPERATOR, and to specify tests by which compliance 186 with the requirements can be verified.

Committee
CTN 209/SC 62
DocumentType
Standard
ISBN
978-2-8322-8647-0
Pages
53
PublisherName
Asociación Española de Normalización
Status
Current

Standards Relationship
BS 4185-5:1971 Identical
IEC 60601-2-20:2020 Identical
EN IEC 60601-2-20:2020 Identical

Access your standards online with a subscription

  • Simple online access to standards, technical information and regulations.

  • Critical updates of standards and customisable alerts and notifications.

  • Multi-user online standards collection: secure, flexible and cost effective.

€89.00
Excluding VAT