BS EN 45502-2-2:2008
Current
The latest, up-to-date edition.
Active implantable medical devices Particular requirements for active implantable medical devices intended to treat tachyarrhythmia (includes implantable defibrillators)
Hardcopy , PDF
English
31-01-2010
1 Scope
2 Normative references
3 Definitions
4 Symbols and abbreviations (optional)
5 General requirements for non-implantable parts
6 Measurement of IMPLANTABLE PULSE GENERATOR and LEAD
characteristics
7 General arrangement of the packaging
8 General MARKINGS for ACTIVE IMPLANTABLE MEDICAL DEVICES
9 MARKINGS on the SALES PACKAGING
10 Construction of the SALES PACKAGING
11 MARKINGS on the STERILE PACK
12 Construction of the NON-REUSABLE PACK
13 MARKINGS on the ACTIVE IMPLANTABLE MEDICAL DEVICE
14 Protection from unintended biological effects being
caused by the ACTIVE IMPLANTABLE MEDICAL DEVICE
15 Protection from HARM to the patient or user caused by
external physical features of the ACTIVE IMPLANTABLE
MEDICAL DEVICE
16 Protection from HARM to the patient caused by electricity
17 Protection from HARM to the patient caused by heat
18 Protection from ionizing radiation released or emitted
from the ACTIVE IMPLANTABLE MEDICAL DEVICE
19 Protection from unintended effects caused by the device
20 Protection of the device from damage caused by external
defibrillators
21 Protection of the device from changes caused by high power
electrical fields applied directly to the patient
22 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
changes caused by miscellaneous medical treatments
23 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
mechanical forces
24 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
damage caused by electrostatic discharge
25 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
damage caused by atmospheric pressure changes
26 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
changes caused by temperature changes
27 Protection of the ACTIVE IMPLANTABLE MEDICAL DEVICE from
electromagnetic nonionizing radiation
28 Accompanying documentation
Annex AA (informative) - Table of cross-references from
90/385/EEC to EN 45502-2-2
Annex BB (informative) - Relationship between the subclauses
of EN 45502-2-2 and the essential requirements
of 90/385/EEC listed in Annex AA
Annex CC (informative) - Notes on EN 45502-2-2
Annex DD (informative) - Code for describing modes of
IMPLANTABLE PULSE GENERATORS
Annex EE (normative) - Interface circuits
Annex FF (informative) - Selection of capacitor C[x]
Annex GG (normative) Calibration of injection network
Test signal of inhibition generator
Annex HH (informative) - Defined terms
Bibliography
Describes requirements that are applicable to IMPLANTABLE CARDIOVERTER DEFIBRILLATORS and the functions of ACTIVE IMPLANTABLE MEDICAL DEVICES intended to treat tachyarrhythmia.
Committee |
CH/150/2
|
DevelopmentNote |
Together with BS EN 45502-2-1, it supersedes BS 6902-1(1990) and BS 6902-1:SUPP1(1996). Supersedes 98/560828 DC. (11/2008)
|
DocumentType |
Standard
|
Pages |
108
|
PublisherName |
British Standards Institution
|
Status |
Current
|
Supersedes |
This Part2-2 of EN45502 specifies requirements that are applicable to implantable cardioverter defibrillators and the functions of active implantable medical devices intended to treat tachyarrhythmia. The tests that are specified in EN45502 are type tests and are to be carried out on samples of a device to show compliance. This part of EN45502 is also applicable to some non-implantable parts and accessories of the devices (see Note1). The characteristics of the implantable pulse generator or lead shall be determined by either the appropriate method detailed in this particular standard or by any other method demonstrated to have accuracy equal to, or better than, the method specified. In the case of dispute, the method detailed in this Particular Standard shall apply. Any aspect of an active implantable medical device intended to treat bradyarrhythmias is covered by EN45502-2-1 Particular requirements for active implantable medical devices intended to treat bradyarrhythmia (Pacemakers). NOTE1 The device that is commonly referred to as an active implantable medical device may in fact be a single device, a combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable parts and accessories if they could affect the safety or performance of the implantable device. NOTE2 The terminology used in this European Standard is intended to be consistent with the terminology of Directive 90/385/EEC. NOTE3 In this European Standard, terms printed in small capital letters are used as defined in Clause3. Where a defined term is used as a qualifier in another term, it is not printed in small capital letters unless the concept thus qualified is also defined. NOTE4 Particular requirements for congestive heart failure devices are under consideration. These types of devices are not covered by this standard.
Standards | Relationship |
EN 45502-2-2 : 2008 COR 2009 | Identical |
EN 45502-2-2 : 2008 COR 2009 | Identical |
EN 45502-2-2:2008 | Equivalent |
AAMI PC69 : 2007 | |
IEC 60068-2-27:2008 | Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock |
EN 60068-2-64:2008 | ENVIRONMENTAL TESTING - PART 2-64: TESTS - TEST FH: VIBRATION, BROADBAND RANDOM AND GUIDANCE |
IEC 60068-2-64:2008 | Environmental testing - Part 2-64: Tests - Test Fh: Vibration, broadband random and guidance |
EN 60068-2-27:2009 | Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock |
IEC 60068-2-47:2005 | Environmental testing - Part 2-47: Test - Mounting of specimens for vibration, impact and similar dynamic tests |
ISO 8601:2004 | Data elements and interchange formats Information interchange Representation of dates and times |
EN 45502-1:2015 | IMPLANTS FOR SURGERY - ACTIVE IMPLANTABLE MEDICAL DEVICES - PART 1: GENERAL REQUIREMENTS FOR SAFETY, MARKING AND FOR INFORMATION TO BE PROVIDED BY THE MANUFACTURER |
ISO 5841-3:2013 | Implants for surgery — Cardiac pacemakers — Part 3: Low-profile connectors (IS-1) for implantable pacemakers |
EN 45502-1 : 2015 | IMPLANTS FOR SURGERY - ACTIVE IMPLANTABLE MEDICAL DEVICES - PART 1: GENERAL REQUIREMENTS FOR SAFETY, MARKING AND FOR INFORMATION TO BE PROVIDED BY THE MANUFACTURER |
EN 28601 : 1992 | DATA ELEMENTS AND INTERCHANGE FORMATS - INFORMATION INTERCHANGE - REPRESENTATION OF DATES AND TIMES |
EN 980:2008 | Symbols for use in the labelling of medical devices |
ISO 11318:2002 | Cardiac defibrillators — Connector assembly DF-1 for implantable defibrillators — Dimensions and test requirements |
EN 60068-2-47:2005 | Environmental testing - Part 2-47: Tests - Mounting of specimens for vibration, impact and similar dynamic tests |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.