EN 14175-4:2004
Current
The latest, up-to-date edition.
Fume cupboards - Part 4: On-site test methods
22-09-2004
Foreword
1 Scope
2 Normative references
3 Terms and definitions
4 General on-site test conditions
5 Commissioning testing of type tested fume cupboards
5.1 Scope and purpose
5.2 Inspections
5.3 Manufacturer's declarations or type test certification
5.4 Face velocity test
5.4.1 Objectives
5.4.2 Test equipment and probe positions
5.4.3 Test procedure and test results
5.4.4 Test report
5.5 Extract volume flow rate test
5.5.1 Objectives
5.5.2 Average face velocity method
5.5.3 Calibrated pressure difference method
5.6 Pressure drop test
5.6.1 Objectives
5.6.2 Test equipment and test procedure
5.6.3 Expression of results
5.7 Air flow visualization
5.7.1 Objectives
5.7.2 Test procedure
5.7.3 Expression of results
5.8 Room air velocity test
5.8.1 Objectives
5.8.2 Test equipment
5.8.3 Probe positions
5.8.4 Test procedure and test results
5.9 Alarm system test
5.9.1 Alarm by overriding the vertical sash stop
5.9.2 Air flow indicator
5.10 Containment test
5.10.1 Objectives
5.10.2 Test equipment, test procedure and data analysis
5.10.3 Test report
5.11 Sound pressure measurement
5.11.1 Objectives
5.11.2 Test procedure
5.11.3 Test report
6 Routine testing of type tested fume cupboards
6.1 Scope and purpose
6.2 Face velocity test
6.2.1 Objectives
6.2.2 Test equipment and test procedure
6.2.3 Test report
6.3 Extract volume flow rate test
6.3.1 Objectives
6.3.2 Sequential average face velocity test
6.4 Pressure drop test
6.5 Air flow visualization
6.6 Alarm system test
6.7 Inspections
7 Qualification testing of non type tested fume cupboards
7.1 Scope and purpose
7.2 Inspections
7.3 Manufacturer's declarations
7.4 Face velocity test
7.5 Extract volume flow rate test
7.6 Pressure drop test
7.7 Air flow visualization
7.8 Room air velocity test
7.9 Alarm system test
7.10 Containment test
7.11 Robustness of containment
7.12 Air exchange efficiency
7.13 Sound pressure measurement
7.14 Illuminance test
8 Routine testing of qualified fume cupboards
9 On-site test reports
9.1 Test report for commissioning testing of type
tested fume cupboards
9.2 Test report for routine testing of type tested
fume cupboards
9.3 Test report for qualification testing of non type
tested fume cupboards
9.4 Test report for routine testing of qualified fume
cupboards
This Part 4 of the European Standard specifies a selection of on-site test methods for the following fume cupboards:- Fume cupboards designed in accordance with Part 2 of this European Standard and type tested in accordance with Part 3 of this European Standard.- Fume cupboards designed in accordance with Part 2 of this European Standard and not type tested.- General purpose fume cupboards designed in accordance with Part 2 of this European Standard which are used under customer-specific conditions, such as for the handling of very toxic substancies or bulky load of the work space.The test methods are designed to be used at the place of installation of the fume cupboard, usually a laboratory. They are used for commissioning after installation, maintenance and qualification purposes.
Committee |
CEN/TC 332
|
DocumentType |
Standard
|
PublisherName |
Comite Europeen de Normalisation
|
Status |
Current
|
Standards | Relationship |
NF EN 14175-4 : 2005 | Identical |
NBN EN 14175-4 : 2004 | Identical |
NEN EN 14175-4 : 2004 | Identical |
NS EN 14175-4 : 1ED 2004 | Identical |
I.S. EN 14175-4:2004 | Identical |
PN EN 14175-4 : 2006 | Identical |
SN EN 14175-4 : 2005 | Identical |
UNI EN 14175-4 : 2005 | Identical |
BS EN 14175-4:2004 | Identical |
UNE-EN 14175-4:2005 | Identical |
DIN EN 14175-4:2004-12 | Identical |
XP CEN/TS 14175-5 : 2010 XP | FUME CUPBOARDS - PART 5: RECOMMENDATIONS FOR INSTALLATION AND MAINTENANCE |
CEN/TS 14175-5:2006 | Fume cupboards - Part 5: Recommendations for installation and maintenance |
BS EN 14175-7:2012 | Fume cupboards Fume cupboards for high heat and acidic load |
I.S. EN 14175-7:2012 | FUME CUPBOARDS - PART 7: FUME CUPBOARDS FOR HIGH HEAT AND ACIDIC LOAD |
UNE-EN 14175-7:2012 | Fume cupboards - Part 7: Fume cupboards for high heat and acidic load |
DIN EN 14175-6:2006-08 | Fume cupboards - Part 6: Variable air volume fume cupboards |
14/30302515 DC : 0 | BS EN 14175-3 - FUME CUPBOARDS - PART 3: TYPE TEST METHODS |
EN 14175-6:2006 | Fume cupboards - Part 6: Variable air volume fume cupboards |
DIN EN 14175-7:2012-08 | FUME CUPBOARDS - PART 7: FUME CUPBOARDS FOR HIGH HEAT AND ACIDIC LOAD |
UNI EN 14175-7 : 2012 | FUME CUPBOARDS - PART 7: FUME CUPBOARDS FOR HIGH HEAT AND ACIDIC LOAD |
EN 14175-7:2012 | Fume cupboards - Part 7: Fume cupboards for high heat and acidic load |
DD CEN/TS 14175-5:2006 | Fume cupboards Recommendations for installation and maintenance |
I.S. CEN/TS 14175-5:2006 | FUME CUPBOARDS - PART 5: RECOMMENDATIONS FOR INSTALLATION AND MAINTENANCE |
BS EN 14175-6:2006 | Fume cupboards Variable air volume fume cupboards |
ISO 11202:2010 | Acoustics Noise emitted by machinery and equipment Determination of emission sound pressure levels at a work station and at other specified positions applying approximate environmental corrections |
BS 2782-1:Method 140E:1982 | Methods of testing plastics. Thermal properties Flammability of a small, inclined test piece exposed to an alcohol flame (laboratory method) |
BS 5588-9:1989 | Fire precautions in the design, construction and use of buildings Code of practice for ventilation and air conditioning ductwork |
EN 14175-1:2003 | Fume cupboards - Part 1: Vocabulary |
BS 4142:1997 | Method for rating industrial noise affecting mixed residential and industrial areas |
BS EN 60079-14:2014 | Explosive atmospheres Electrical installations design, selection and erection |
EN 61672-1:2013 | Electroacoustics - Sound level meters - Part 1: Specifications |
ISO 5221:1984 | Air distribution and air diffusion Rules to methods of measuring air flow rate in an air handling duct |
EN 14175-3:2003 | Fume cupboards - Part 3: Type test methods |
EN ISO 11202:2010 | Acoustics - Noise emitted by machinery and equipment - Determination of emission sound pressure levels at a work station and at other specified positions applying approximate environmental corrections (ISO 11202:2010) |
EN 14175-2:2003 | Fume cupboards - Part 2: Safety and performance requirements |
Access your standards online with a subscription
Features
-
Simple online access to standards, technical information and regulations.
-
Critical updates of standards and customisable alerts and notifications.
-
Multi-user online standards collection: secure, flexible and cost effective.